- Aswin L. HoffmannTechnische Universität Dresden, OncoRay—National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus#Technische Universität Dresden, Department of Radiotherapy and Radiation Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus
- Titel
- Image Performance Characterization of an In-Beam Low-Field Magnetic Resonance Imaging System During Static Proton Beam Irradiation
- Zitierfähige Url:
- https://nbn-resolving.org/urn:nbn:de:bsz:14-qucosa2-830058
- Quellenangabe
- IEEE Transactions on Radiation and Plasma Medical Sciences
Erscheinungsjahr: 2022
Jahrgang: 6
Heft: 3
Seiten: 271-281
E-ISSN: 2469-7303 - Erstveröffentlichung
- 2022
- Abstract (EN)
- Image guidance using in-beam real-time magnetic resonance (MR) imaging is expected to improve the targeting accuracy of proton therapy for moving tumors, by reducing treatment margins, detecting interfractional and intrafractional anatomical changes and enabling beam gating. The aim of this study is to quantitatively characterize the static magnetic field and image quality of a 0.22T open MR scanner that has been integrated with a static proton research beamline. The magnetic field and image quality studies are performed using high-precision magnetometry and standardized diagnostic image quality assessment protocols, respectively. The magnetic field homogeneity was found to be typical of the scanner used (98ppm). Operation of the beamline magnets changed the central resonance frequency and magnetic field homogeneity by a maximum of 16Hz and 3ppm, respectively. It was shown that the in-beam MR scanner features sufficient image quality and influences of simultaneous irradiation on the images are restricted to a small sequence-dependent image translation (0.1–0.7mm) and a minor reduction in signal-to-noise ratio (1.3%–5.6%). Nevertheless, specific measures have to be taken to minimize these effects in order to achieve accurate and reproducible imaging which is required for a future clinical application of MR integrated proton therapy.
- Andere Ausgabe
- Link zum Artikel der zuerst in der Zeitschrift 'IEEE Transactions on Radiation and Plasma Medical Sciences' bei IEEExplore erschienen ist.
DOI: 10.1109/TRPMS.2021.3085991 - Freie Schlagwörter (DE)
- Bildqualität, Magnetresonanz (MR) geführte Strahlentherapie, Magnetometrie, MRT, Protonentherapie (PT)
- Freie Schlagwörter (EN)
- Image quality, magnetic resonance (MR) guided radiotherapy, magnetometry, MRI, proton therapy (PT)
- Klassifikation (DDC)
- 621.3
- 624
- Verlag
- IEEE, New York
- Version / Begutachtungsstatus
- angenommene Version / Postprint / Autorenversion
- URN Qucosa
- urn:nbn:de:bsz:14-qucosa2-830058
- Veröffentlichungsdatum Qucosa
- 20.01.2023
- Dokumenttyp
- Artikel
- Sprache des Dokumentes
- Englisch
- Lizenz / Rechtehinweis